Plinski analizator MAT1500
MAT1500 gas analyser for O₂ and CO₂ measurement

Plinski analizator MAT1500
Premium model with a ZrO₂ sensor for O₂ and an NDIR sensor for CO₂: measurement takes about 10 s at most, and min/max detection delivers precise results even with a small sample volume.
Technical data
- Model
- MAT1500
- O₂ sensor
- ZrO₂ sensor, 0–100 vol.-%, accuracy ±0.02 vol.-% abs. (< 1 vol.-%) / ±3 vol.-% rel. (> 1 vol.-%), resolution 0.001 vol.-%, drift-free
- CO₂ sensor
- two-channel NDIR, 0–100 vol.-%, repeatability ±0.5 vol.-%, resolution 0.1 vol.-%, drift < 1 %/month
- Scales
- O₂, CO₂, N₂
- Measurement time
- 10 seconds
- Minimum headspace volume
- 5 ml
- Adjustment
- automatic (menu-driven)
- Methods & users
- over 10,000
- Operation
- 7.0″ capacitive touchscreen 800 × 480 (keyboard, mouse, barcode scanner optional)
- Interfaces
- 1× USB, 1× RS-232, 1× Ethernet, LIMS
- Additional
- input of product and batch designations, alarm limits, user management, measured value stabilisation, air pressure display
- Ambient temperature
- 10–50 °C
- Dimensions (W × H × D)
- 200 × 150 × 350 mm
- Weight
- 4.6 kg
- Operating voltage
- 100–250 V AC, 50/60 Hz, max. 60 W
Similar products
Interested in this product?
Send an enquiry and we will respond with a tailored quotation or advice.